GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm Page no. 177 Chapter 16
S91 Need user action
on AWS
Operator input is required on
AWS.
Respond to the AWS pop-up.
S92 AWS busy, please
wait...
The AWS is busy, and the current
operation may take some time.
No action required, wait for completion of the cur-
rent operation.
S93 AWS busy The AWS is busy (the wait should
be short).
No action required, wait for completion of the cur-
rent operation.
S94 System performing
test, wait
An application is running on the
system and performing tests that
do not allow acquisition (example
auto-diagnostics).
Wait a few minutes.
S95 Medical Applica-
tion not active
AWS has not yet entered applica-
tion mode. This message is
always displayed after system
boot and after each close exam
even during review.
It is displayed because no appli-
cation is ready to get the image.
Start a new exam or QAP acquisition.
S96 System activity
conflict
A diagnostic or calibration appli-
cation is running while the medi-
cal application is active.
Check if several applications are running. If yes,
close one.
S97 AWS reset in prog-
ress, wait
AWS is being reconfigured. The
message is displayed until recon-
figuration is complete.
No action required.
S98 Exp. inhibited CAL/
IQ in progress
AWS is being reconfigured. The
message is displayed until recon-
figuration is complete.
No action required.
S99 OFFSET ACQUISI-
TION IN PROGRESS
AWS is being reconfigured. The
message is displayed until recon-
figuration is complete.
No action required.
Message on
X-ray Console
Explanation Corrective Action