Error messages
16-errors.fm Page no. 185 Chapter 16
A05 Compression
upper position is
reached
1) Compression release move-
ment is stopped because the
upper position with the current
paddle is reached.
or
2) The upper position of the pad-
dle holder is reached.
1) If you have a paddle, remove it to go to the pad-
dle holder upper position.
or
2) If the paddle is removed and the paddle holder
is not at upper position, note ALL messages and
contact GEMS Field Service.
A06 Lift end position is
reached
Lift up or down movement is
stopped because the upper or
lower position is reached.
Move the lift in the opposite way.
A07 Lift is not opera-
tional
Lift movement is not allowed
because the lift device is not oper-
ational.
None
A08 Max manual com-
pression force is
reached
The system informs the operator
that the maximum force allowed is
reached.
No more compression is available, even using
manual compression.
A09 Max power driven
compression force is
reached
The system informs the operator
that the maximum force allowed
for motorized movement is
reached.
1) Use the manual control to compress more.
2) If the maximum power driven force is < 20daN
use the Medical Menu to set a higher value.
A10 Motion stopped.
Multiple movements
are requested
The current movement is stopped
because another non-compatible
movement is requested:
This advisory message occurs:
A. If you press a user interface
(button or footswitch) during a
pre-set movement.
B. If you press a pre-set button
while a previous movement is in
progress (including deceleration
phase).
Release all footswitches or buttons and wait for all
movement to stop before pressing on the foot-
switch or the gantry control movement button
again.
A12 Move tube head to
center position
Rotation movement has been
requested with the tube head
angled away from the center posi-
tion (0°).
1) If the Stereotactic Positioner is installed, switch
back to angulation mode and move tube head to
the center position (0°).
2) If the Stereotactic Positioner is not installed, an
error has occurred:
A. Power OFF/ON from the X-ray Console.
B. If the problem persists, note ALL messages
and contact GEMS Field Service.
A14 Risk of collision
with floor
Lift or Rotation movement is
stopped to avoid a collision with
the floor.
Select another incidence.
Message on
Gantry Readout
Explanation Corrective Action