EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 16 Page no. 190 16-errors.fm
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
H23 Rotation brake fail-
ure
The system is set in emergency
mode:
- The arm rotation brake has
failed; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H24 Rotation software
error
The system is set in emergency
mode:
- An error occurred on the rotation
device CPU; the system is set in
"halt" state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H25 Halt. Rotation
Uncontrolled
The system is set in emergency
mode:
- Uncontrolled Rotation move-
ment; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H27 Compression
brake failure
The system is set in emergency
mode:
- The compression device is in
fault state; the system is set in
"halt" state.
-It is not possible to perform any
movement or image acquisition.
1) Power OFF/ON from the X-ray Console.
2) Try to perform compression motion. If the prob-
lem persists, note ALL messages and contact
GEMS Field Service.
H28 Control button fail-
ure on arm
The system is set in emergency
mode:
- An arm movement control button
on an arm keypad (behind the
image receptor, left or right) has
failed; the system is set in "halt"
state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
H29 Control button fail-
ure on head
The system is set in emergency
mode:
- An arm movement control button
on an X-ray tube head keypad
(left or right) has failed; the sys-
tem is set in "halt" state.
-It is not possible to perform any
movement or image acquisition.
1) Decompress the patient, using the manual com-
pression knobs.
2) Power OFF/ON from the X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
Gantry Readout
Explanation Corrective Action
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Related product manuals