GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Error messages
16-errors.fm Page no. 195 Chapter 16
X04 Collimation com-
munication failure
The Field Of View cannot be set
at the correct format:
- The Collimator is not communi-
cating.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X05 Collimation control
failure
The Field Of View cannot be set
at the correct format:
- The Collimation function is
failed.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X06 Generator commu-
nication failure
The Gantry state cannot allow an
acquisition:
- The Gantry is not communicat-
ing with the Generator.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X07 Invalid bucky type The Gantry does not recognize
the Bucky device.
- It is not possible to perform an
image acquisition.
1) Unplug the Bucky.
2) Reinstall the Bucky: check that it is correctly
installed and switched on (initialization noise
heard).
3) Power OFF/ON from the X-ray Console.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
X09 Compression pad-
dle detection error
The Gantry does not recognize
the paddle.
- The compression thickness can-
not be computed accurately.
- It is not possible to continue the
current exam, exposure is inhib-
ited.
1) Remove and reinstall the paddle.
2) If the problem persists, power OFF/ON from the
X-ray Console.
3) If the problem persists, note ALL messages and
contact GEMS Field Service.
X10 X-ray beam form
software error
An error occurred during X-ray
beam format computation by the
Gantry CPU.
- It is not possible to perform an
image acquisition.
1) Power OFF/ON from the X-ray Console.
2) If the problem persists, note ALL messages and
contact GEMS Field Service.
X11 Tube head com-
munication failure
The tube focal spot size cannot be
set:
- The tube head device is not
communicating.
- It is not possible to perform an
image acquisition.
1) If you hear the tube tilt motor, the system is try-
ing to recover itself.
2) If the message "S61 X-ray beam not ready" is
present more than 60s on the x-ray control con-
sole, power OFF/ON from the X-ray console.
3) If the problem persists, shut down the system
from the Browser.
4) If the problem persists, note ALL messages and
contact GEMS Field Service.
Message on
Gantry Readout
Explanation Corrective Action
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals