Chapter 4 Page no. 38 4-cons.fm
GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
X-ray Console
3-1 Access to setup menus
DEUTSENGLFRANC PORTUITALIESPAN
• With the X-ray Console in normal application mode, press the Setup button to enter the "menu tree"
shown below.
• To access a function, press the button below the displayed name of the function.
• To back out of a function (move "up the tree"), press the Setup button again.
view medical language PASSWORD
DECOMP COMP LIFT BEEP ROTATE STEREO
+ - VALID
FAST SLOW
YES NO
recumbent
medical set up
DECOMPRESSION HEIGHT: 21 cm
END OF EXPOSURE DECOMP:
DECOMP HEIGHT LIGHT
MAX COMPRESSION FORCE: 16 daN
FORCE SPEED
PRESET SPEED
FAST SLOW
FAST SLOW
COMPRESSION SPEED: FAST
DEFAULT OBLIQUE ANGLE: 45 deg
DEFAULT LIGHT DURATION: 3 min
LIFT FOOTSWITCH SPEED: SLOW
WITH BEEP: YES
ROTATION SPEED: SLOW
+ - VALID
+ - VALID
+ - VALID
YES NO
+ - VALID
Function available for service only
FOR TRAINING PURPOSES ONLY!