Xcellerex
™
XDM 50-1000, XDUO
100-1000 mixers
Operating Instructions
Original instructions
Translation disc enclosed

Table of Contents

Questions and Answers

  • J
    Janet RothSep 9, 2025
    What to do if an error code is shown on the VFD display of GE Xcellerex XDM 200 Mixer?
    • C
      Corey LeeSep 9, 2025
      Refer to the VFD manufacturer's manual for error code clarification. If the problem cannot be solved on site, contact your GE representative for help.
  • B
    Brian FloresSep 8, 2025
    What to do if different probes in the same solution show different pH readings on GE Xcellerex XDM 200 Mixer?
    • J
      Jason BrownSep 8, 2025
      If different probes in the same solution show different pH readings: * Calibrate pH probes before use, because one of the pH probes may not be calibrated. * Replace the probe, because the probe may be at the end of its life cycle.
  • D
    Dana CastilloAug 22, 2025
    What to do if pH is not coming to equilibrium on GE Mixer?
    • R
      rachel99Aug 22, 2025
      If the pH is not reaching equilibrium even when the pumps are running, consider these possible causes and solutions: * Ensure the deadband is not set to 0, as pH control requires a deadband to prevent alternating introduction of acid and base. Define an appropriate deadband value for your process. * Check if the acid/base supply is empty and fill it if necessary. * Prime the pumps to ensure they are functioning correctly. * Inspect the feed lines for kinks or clamps and free them to ensure proper flow. * Conduct a study to verify the control parameters, considering factors like media quantity, pump type, acid and base concentration, mapping strategy, media buffering capability, and mixing of vessel contents.
  • A
    Angel SantosSep 5, 2025
    What to do if tubing is worn out on GE Xcellerex XDM 200?
    • C
      Chelsey ChristensenSep 5, 2025
      All tubing used in the pumps will wear out with use over time. Move the section of worn out tubing to the left or right. Use only tubing rated for use in peristaltic pump head.
  • C
    Crystal DawsonSep 4, 2025
    What to do if the pump is not turning on GE Xcellerex XDM 200 Mixer?
    • P
      Penny LewisSep 4, 2025
      If the pump is not turning: * Verify that the cause of the E-STOP condition is eliminated, and clear the E-STOP. Try to use the pumps again. * Clear the interlock condition.
  • G
    Gary SimsSep 3, 2025
    What to do if pH sensor changes are very slow on GE Mixer?
    • M
      Michele WilsonSep 3, 2025
      If pH sensor changes are very slow, and two sensors change their readings at different rates, the pH sensor may need cleaning. If available, switch the input to the second pH channel. Keep probes clean between uses, and clean them with a solution designed to maintain their properties. Replace probes regularly.
  • A
    Alexis HayesSep 1, 2025
    What to do if temperature monitoring is not working as expected on GE Mixer?
    • W
      Wendy BurtonSep 1, 2025
      If temperature monitoring is not working as expected: * Switch to the second input of the temperature probe if possible. If the second input is not present or is not working, replace the probe. * Ensure the temperature probe is inserted into the thermowell.
  • M
    Miguel JacksonAug 30, 2025
    What to do if the agitator does not start on GE Xcellerex XDM 200 Mixer?
    • A
      Amber DeleonAug 30, 2025
      If the agitator does not start, consider the following: * Check if the Min Weight interlock is active. Check the weight of the disposable bag. * If the system is in the state of E- STOP, tap RESET on the local control software home screen and start the agitator when the system has been restarted after emergency shutdown. Note: The agitator cannot be started while the system is in the state E-STOP.
  • M
    Mrs. Tammy MitchellAug 29, 2025
    What to do if agitator motor cannot be started on GE Xcellerex XDM 200?
    • J
      Jessica SmithAug 29, 2025
      If the agitator motor cannot be started, check the following: * Ensure the power switch on the I/O Cabinet is in the I ON position. * Verify the power cable is connected. * If the issue persists, the drive may be broken, and you should contact your GE representative to replace it.
  • A
    Ashley Rodriguez PhDAug 27, 2025
    Why is variable frequency drive (VFD) not functioning as expected in GE Xcellerex XDM 200 Mixer?
    • J
      Jamie ChaneyAug 27, 2025
      The variable frequency drive (VFD) might not be functioning as expected due to: * The motor being overloaded. Acknowledge the agitator fault and attempt to restart the motor. If the error persists, contact GE for service. * The power cable not being connected. Connect the power cable. * The drive is broken. Contact your GE representative to replace the drive.

Summary

Safety Instructions for Xcellerex Mixers

General Safety Precautions

Covers essential safety measures including hazard awareness, proper operation, and personnel qualification.

Emergency Procedures

Details procedures for emergency shutdowns, power failures, and system restarts for XDM and XDUO mixers.

Xcellerex XDUO Mixer Interlocks

Explains the interlock functions in the XDUO mixer that prevent operation under specific unsafe conditions.

System Description for Xcellerex Mixers

Safety Hardware Features

Explains the integrated safety systems such as emergency stop, labeling, burst discs, interlocks, and fuses.

Mixer Automation and Control

Describes how XDM and XDUO mixers are automated and controlled, including VFD and local control software.

User Interface for XDUO Mixer

Alarm Management Screens

Explains how to view active alarms and configure alarm limits for different system parameters.

Detail Screens for Process Variables

Describes the detail screens for agitator, conductivity, pH, temperature, and weight, including navigation buttons.

Installation Procedures for Xcellerex Mixers

Safety Precautions During Installation

Highlights essential safety warnings and precautions to be followed during the installation process.

Mixer Installation Steps

Details the procedures for moving the unpacked mixer, electrical installation, and configuring local control software.

Preparation for Xcellerex Mixer Operation

Safety Precautions Before Preparation

Covers critical safety warnings related to PPE, magnets, high voltage, and general handling during preparation.

Preparing the Disposable Bag

Details the process of unpacking and preparing the disposable bag assembly for installation.

Preparing and Installing Probes

Guides on preparing, calibrating, and installing pH and conductivity probes, including autoclaving.

Operation of Xcellerex Mixers

Safety Precautions During Operation

Highlights safety measures for operation, including handling hazardous substances, ventilation, and hot surfaces.

Operating the Xcellerex XDM Mixer

Provides instructions for starting the system, operating the agitator motor, and monitoring key parameters.

Operating the Xcellerex XDUO Mixer

Details how to start the system, log in/out, connect to FlexFactory, and manage data logging.

Adding Liquid Contents to the Bag

Guides on adding liquids into the disposable bag, including preparation and filling procedures.

Adding Solid Contents to the Bag

Explains how to add solid materials into the disposable bag, considering pre-addition aspects and procedures.

Finishing a Batch Run

Covers the final steps of a batch run, including collecting product, removing the bag, and shutting down the system.

Maintenance of Xcellerex Mixers

Safety Precautions for Maintenance

Emphasizes safety measures for maintenance personnel, including trained staff, power off, and contamination hazards.

Lock-Out; Tag-Out (LOTO) Procedures

Provides detailed instructions for both turning off (LOTO) and turning on (LOTO) the instrument safely.

Troubleshooting Xcellerex Mixers

GE Xcellerex XDM 200 Specifications

General IconGeneral
ModelXcellerex XDM 200
Capacity/Nominal Volume200 L
Volume Range50-200 L
Mixing SpeedVariable
Vessel MaterialStainless steel
Single-Use LinerYes
Liner MaterialPolyethylene
MaterialStainless steel and polyethylene
SterilizationGamma irradiation for single-use components
Control SystemIntegrated control system
Power SupplyElectrical
TypeSingle-Use Mixer