Genevac HT-8 & HT-12 Series II Evaporating Systems
04-4541 Issue 1-9 – June 2008 Page 51 of 64
General
Suspected leak - pressure of no better than 5
mbar.
Condenser
Condenser not getting to temperature ( - 30° C ).
Electrical
No power to any part of the system.
No power to the pump and condenser.
No power to the chamber.
Door will not open - system powered up.
Probable cause
If the tube is satisfactory, reconnect the tube to
the condenser and disconnect the heated inlet
from the chamber at the condenser port.
Connect the vacuum transducer to the
condenser port and switch the pump on.
If the pressure is no better than 1mbar, contact
Genevac Service for assistance.
If the pressure is better than 1mbar, reconnect
the chamber and switch the pump on.
If the same conditions prevail, contact Genevac
Service for assistance.
Probable cause
Contact Genevac Service for assistance.
Probable cause
Check the circuit breaker on the mains supply to
the system and reset if necessary.
Check switches.
Check leads (pump to evaporator).
If the fault persists, contact Genevac Service for
assistance.
As above
As above
Contact Genevac Service for assistance.
If it is imperative that you retrieve your samples
immediately, remove the door pawls.
Access points above and below the door, allow
the door pawls to be disconnected.
Plan view of upper
access - lower access
below the door-use a 5
mm Allen key.