Genevac HT-8 & HT-12 Series II Evaporating Systems
Page 62 of 64 04-4541 Issue 1-9 – June 2008
Appendix C
Programming guide for mixtures of high and low boiling point solvents
Drying High - Low Mixtures
Run number
1 2 3
Run Data Locked
N N N
Next Run Number
2 3 0
Run Name
High - Low Mix High - Low Mix part 2 High - Low Mix part 3
Sample Holder Type
Your Holder Your Holder Your Holder
Sample Volume ml
Your Volume Your Volume Your Volume
Solvent DMSO or DMF & DCM DMSO or DMF & DCM DMSO or DMF & DCM
SampleGuard Control Temp C
40 40 40
SampleGuard Control Channel
1 1 1
CoolHeat Enable Pressure
50 50 50
Chamber Temperature control
preheat to 15 preheat to 40 preheat to 30
Rotor Speed
High High Low
Heat-Off Option
Elapsed Time Heat Flow Elapsed Time
Heat-Off Elapsed Time
ask for time - 00:00
End-of-Run Option
Elapsed Time - Elapsed Time
End-of-Run Elapsed Time
ask for time - 01:00
Auto Defrost & Drain
Y N Y
Pressure Control Regime
Variable Dri-Pure Full Vacuum Full Vacuum
Dri-Pure Start Pressure (mbar)
750 - -
Dri-Pure End Pressure (mbar)
70 - -
Dri-Pure Ramp Time 55 - -
Final Control Pressure
65 - -
Notes
Remove MeCN Dry high BP solvent Final Drying
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents