Pressed for less than one second it scrolls the possible selections.
Press for less than one second to exit from the selection function.
Pressed for less than one second it confirms the showed selection.
Once entered in configuration mode we have access, in sequence, to the three main
menus:
UNIT 2 : to enable the alternative units (oz, lb, ct, Pcs, %)
SETUP : to modify the transmission parameters, to enable or disable
functions, to store personalised configurations.
CALIBRATION : to start the calibration process.
- NOTE -
▪ If no key is pressed while in the configuration menu, the menu moves back to a
previous level every 20 seconds until exiting from the configuration mode. The
balance goes back automatically to operating mode, without modifying any
parameters.
▪ At section 6.4 there is a short graphic form of the complete menu’s tree, of the
submenus and of the available parameters.
ATTENTION !
If the balance stops working because of incorrect operations on the keyboard, operate
as follows:
c- unplug the power supply unit from the mains socket and wait some seconds
d- plug in again the power supply unit and simultaneously press the ON-OFF key.
Release the key when the display shows “EU-C”.
For the version “BP” (models with rechargeable battery) operate as follows:
d- turn off the power switch put on the rear of the balance and wait some seconds;
e- turn on the power switch and, simultaneously, press the ON-OFF key of the
keyboard.
f- Release the ON-OFF key when the message “EU-C” appears.
In both cases the message "SEt Factory" will confirm the reset of default parameters
(factory settings).
6.1 “UNIT 2” MENU
This is the first menu visualized once entered in configuration modality. Through this menu
it’s possible select and enable a second alternative unit, operating as follow:
▪ When the display shows “UNIT 2”, press for less than one second the ON-OFF key
to enter the submenu.