To transmit data to a peripheral that can
not send specific commands to the
balance.
The continue transmission is activated
at power up, so that the transmission of
the measured data follows every
measure carried out by the balance.
(Default) The balance transmits a data
just once when a valid command is
issued from serial port.
Data stable transmission:
▪ With the EUROPE protocol the
balance transmits only once, as
soon as a valid command is sent,
but only if the data of the balance is
stable.
▪ With the CRYSTAL protocol the
balance transmits like the
EUROPE protocol but, in presence
of an unstable data, it will carry out
however a transmission after 15
second. This is in order to not
leave the system blocked.
(Default) The balance transmits in
whichever condition.
Disables the CALIBRATION function.
(Default) Enables the CALIBRATION function.
(Default) Enable the automatic zero tracking. Useful to
cancel small drift of the zero reading.
Disable the automatic zero tracking.
(Default) Set the filter with slow response applicable to
“disturbed” environments or for high accuracy weighing.
Set the filter at medium time response.
Set the filter with fast response. Applicable to quickly
changing weights (ex. for dosing).
Enables the automatic switch-off of the display when the
balance is unutilised for more than 10 minutes.
(Default) Disables the automatic switch-off of the display