13 Continuity Testing for Diode Tests
If the “acoustic signal” function has been activated, the
instrument generates a continuous acoustic signal exclu-
sively in the measuring range 0 ... 1.8 V (3¾ place display)
within a measurement value range of 0 to approx. 0.1 V.
The limit value can be selected in the “SEt” menu mode:
SEt triG cont i cont i V XXX mV .
Activate continuity testing (acoustic signal ON):
Briefly press the multifunction key ESC|FUNC.
Activation is acknowledged with an acoustic signal.
The symbol is simultaneously displayed at the LCD.
OL is displayed if the DUT is not connected.
Connect the measurement cables to the DUT.
The signal generator function is activated by pressing the
ESC|FUNC multifunction key once again.
14 Signal Generator
Individual pulses or pulse packets with an amplitude of
approx. 3 V and a frequency of 1 to 1000 Hz can be gen-
erated with the help of the signal generator function.
Activate the signal generator parameter (see below).
Turn the function selector switch to the “ ” position.
Connect the DUT to the” and “V” sockets with
measurement cables.
Press the ESC|FUNC multifunction key twice.
“Hz” appears at the display, as well as the pulse fre-
quency entered to the “Setup” menu.
Select the desired frequency with the help of the
DATA|CLEAR key. The following values are available:
1, 2 , 5, 10, 20, 50, 100, 200, 500, 1000 [Hz] or USEr
Start the signal generator.
Continuous operation: If rEPEt has been selected, the sig-
nal generator is started automatically
Single sequence: If onCE has been selected, press the
MAN|AUTO key to start each individual sequence.
Setting Parameters
Enter the “SEt” menu mode (see flow chart on following
page as well as chapter 21).
USEr Frequency
SEt GEnEr USEr or xxxx Hz 
USEr = 1 ... 999 Hz, step size 1 Hz
xxxx = 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000 [Hz]
count – Number of Pulses per Pulse Packet
count: number of pulses: 1...99999
onCE/rEPEt – Single Sequence/Continuous Operation
onCE: single sequence
rEPEt: continuous operation
dELAY – Interval Between 2 Pulse Packets
dELAY: pause of 1 to 99 999 ms
dELAY: no function if onCE has been selected
Duty cycle: approx. 50%