32 GMC-I Gossen-Metrawatt GmbH
triG = out
Measurement values are stored to memory if the following
prerequisite has been fulfilled: At least one measurement
value occurs within the limits defined by H.triG and L.triG,
and one of the subsequent measurement values must vio-
late these limits.
SEt triG V  out H.triG ...
triG = in
Measurement values are stored to memory if the following
prerequisite has been fulfilled: At least one measurement
value occurs which violates the limits defined by H.triG and
L.triG, and one of the subsequent measurement values is
within these limits.
SEt triG V  in H.triG ...
triG = St-ou
Only those measurement values are stored to memory,
which do not lie within the limits for H.triG and L.triG.
SEt triG V  St-ou H.triG ...
triG = St-in
Only those measurement values are stored to memory,
which do lie within the limits for H.triG and L.triG.
SEt triG V  St-in H.triG ...
20.2.1 Trigger Function Parameters
H.triG/L.triG – Upper Limit/Lower Limit
Each trigger function is assigned its own upper and lower
limits as trigger parameters. The upper limit should be
greater than the respective lower limit.
The predetermined trigger threshold is evaluated according
to the selected measuring range, independent of whether
or not automatic measuring range selection is activated.
Entering upper and lower trigger thresholds in digits:
SEt triG V in
H.triG XXXXX 
L.triG XXXXX .
Note!
Values for H.triG and L.triG are also used as trigger
parameters for cont i or EVENTS. The table in chap-
ter 18.1 contains examples for the entry of trigger
thresholds in digits.