GMC-I Gossen-Metrawatt GmbH 41
GB
21.2 Description o Parameters in the inFo Menu
tiME – Time Setting
SEt inFo ↵ tiME ↵ 10:24 ↵ 10:24:42 (hh:mm:ss) ↵
21.01 ↵ 21.01.99 (TT.MM.JJ)
OCCUP – Memory Occupancy (METRA HIT 22M/26M/MIL only)
See description: chapter 20, page 28.
t.rAM (RAM Test) – Testing RAM (METRA HIT 22M/26M/MIL only)
Attention!
!
Activating this function deletes all stored measure-
ment values from memory.
Do not perform the RAM test while any of the follow-
ing functions are active:
events counter or memory mode.
Starting the RAM test:
SEt info ↵ t.rAM ↵ no YES ↵
No other functions may be activated during the RAM test
(the “bUSY” message is displayed). The test lasts approxi-
mately 1 minute. Two test samples are written to memory,
and are subsequently read out.
If the test is completed successfully, “PASSE” appears at
the display.
Significance of possible messages:
bUSY RAM test is running
PASS Test successfully completed
Err1 Test sample for this test is faulty
Err2 Test sample for a previous test is faulty
If Err1 and/or Err2 occur, a hardware problem may exist.
Send the multimeter to our Repair and Replacement Parts
Service Department.
bAtt – Battery Voltage
SEt inFo ↵ bAtt ↵ 3.0 V.
21.3 Default Settings
Previously selected settings can be deleted, and default
settings can be restored. This may be helpful in the follow-
ing situations:
• After the occurrence of hardware or software problems
• If you feel that the multimeter is not functioning properly