A/D default: No response
from the A/D converter.
Xtrem ADPD module damaged.
Load cell input signal too high
(>20mV).
Check the scale for some load
cell / wiring default.
Load cell input signal too low
(<-20mV).
Load cell input signal out of
range (>30mV)
Load cell excitation and A/D
converter switched off
because a short-circuit.
Check wiring and load cells.
Data flash memory storage is
corrupted and cannot be reset
to the factory defaults.
Break the verification seal to
open the XTREM cover, set to
“unlock” the sealing switch. The
instrument must be adjusted
and verified / calibrated again.
Calibration weight > Max.
Adjust weight should be < Max.
The division should be >10
A/D counts.
Resolution is too high. Change
the division to a higher value.
A stable measurement cannot
not be obtained for adjusting
the scale.
Increase the filter level.
Overload:
Weight > Max+9·div
Settings value cannot be
changed because the “sealing
switch” is at lock position.
Legally relevant settings cannot
be changed without breaking
the verification seal.
Negative weight (weight < -
19e)
Initial zero setting in progress