Specifications
4-5
4.2 Function Specifications
Function Specifications
Trigger Functions
Item Description
Display screen Waveform screen + Digital screen, Expanded Waveform screen,
Digital screen + Calculation Display screen
Note: Can be key-toggled
Sampling interval 100 ms/10 ch maximum
100, 200, 500 ms; 1, 2, 5, 10, 20, 30 s; 1, 2, 5, 10, 20, 30 min; 1 h
EU (scaling function) 4 points can be set for each channel
Review function Data replay during data capture
Data save functions Capture to internal memory
Capture to USB memory
The setup data can be saved (to internal or USB memory)
Copy of data screen can be saved (to internal or USB memory)
Calculation Types of statistical calculation : Average value, peak value,
maximum value, minimum value, RMS
Number of operations : maximum of 2 can be set simultaneously
Method : Realtime and between cursors specified (during data replay)
Note: Calculation results are displayed in Digital screen + Calculation Display screen
Search functions Function: Search the captured data for the required number of points
Search type: Channel Pulse, Logic, Level, Alarm search
Annotation input function Function: A comment can be input for each channel
Inputtable characters: Alphanumerics
Number of characters: 11 (displayed up to 8 characters)
Item Description
Repeat Trigger Off, On
Trigger types Start: Data capture starts when a trigger is generated.
Stop: Data capture stops when a trigger is generated.
Trigger conditions Start: Off, Level, External, Time, Alarm
Stop: Off, Level, External, Time, Duration, Alarm
Alarm judgment modes Analog : H (
), L (), Window In, Window Out
Logic : H (
), L ()
Pulse : H (), L (), Window In, Window Out
Channel combination OR, AND