In order to satisfy the nitrogen conditions generated by the Nitrogen Generator, the air
compressor must satisfy the following conditions.
Pressure : 7kgf/cm
2
(0.7Mpa) or more, Recommended : 8kgf/cm
2
or more
Flow rate : 100L/min or more
If the air flowing into the nitrogen generator does not receive sufficient pressure, it will not
proceed when selecting Nitrogen curing.
In an environment where air lines are branched instead of using an air compressor alone,
sufficient nitrogen may not be injected during nitrogen hardening due to excessive use of air
from other devices.
If there is excessive moisture in the compressed air flowing into the nitrogen generator, the
nitrogen generator may be clogged and the nitrogen purity may decrease. To prevent moisture
from accumulating in the air filter, put a tissue on the drain part and press the drain part to
remove moisture. If possible, use of an air dryer is recommended..