If you want to proceed with the next curing immediately after curing is finished, the internal
temperature is high and the curing conditions may change slightly.
Use it after having enough cool-time or cooling the inside for a while using an air gun.
After curing is complete, clean the tray plate before use. The tray plate is a consumable
part, and it is recommended to replace it with a new part if the contamination of the bottom
surface has progressed to the point where it cannot be cleaned after long-term use.
□ Nitrogen curing (necessary to connect nitrogen generator)
When performing nitrogen hardening, make sure that the nitrogen hardening selection is
activated.
After selecting nitrogen curing, start curing after setting the same as normal curing.
When curing starts, nitrogen injection starts first with the sound of the solenoid valve opening,
and the progress page displays a graph and remaining time for nitrogen injection.
If the nitrogen ratio in the curing chamber reaches the set value within the time for pre-
nitrogen injection (default value is 5 minutes), the graph and remaining time are changed to
the curing time and restart.
Please read the precautions (p12) when installing the nitrogen generator.