GDS-800 Series Digital Storage Oscilloscope Operation Manual 
 
69
When the oscilloscope is set up with a RS232 interface, please check the following 
points: 
z  Do not connect the output line of one DTE device to the output line of the 
other. 
z  Many devices require a constant high signal on one or more input pins. 
z  Ensure that the signal ground of the equipment is connected to the signal 
ground of the external device. 
z  Ensure that the chassis ground of the equipment is connected to the 
chassis ground of the external device. 
z  Do not use more than 15m of cable to connect devices to a PC. 
z  Ensure the same configurations are used on the device as the one used on 
PC terminal. 
z  Ensure the connector for the both side of cable and the internal connected 
line are met the demand of the instrument. 
Computer’s Connection 
A personal computer with a COM port is the essential facilities in order to operate 
the digitizing oscilloscope via RS232 interface. 
The connections between oscilloscope and computer are as follows: 
I.  Connect one end of a RS232 cable to the computer. 
II.  Connect the other end of the cable to the RS232 port on the oscilloscope. 
III.  Turn on the oscilloscope. 
IV.  Turn on the computer.