OPERATION 
25 
CH3 Independent Mode 
 
Background / 
Connection 
The CH3 rating is 2.5V/3.3V/5V, 3A fixed. It 
works independently from CH1 and CH2, 
regardless of their modes. 
 
Output rating 
2.5V/3.3V/5V, 3A fixed 
No Tracking 
Series/Parallel 
CH3 does not have tracking series/parallel mode. 
Also, CH3 output is not affected by CH1 and CH2 
modes. 
Panel operation 
1.  Connect the load to the front 
panel CH3 +/− terminal. 
(the diagram shows non-
European terminals) 
 
2.  Select the output voltage, 
2.5V/3.3V/5V using the 
CH3 voltage selector key. 
 
    GPD-3303 Series User Manual 
26 
 
3.  To turn on the output, press 
the output key. The key 
LED turns on. 
 
CV → CC 
When the output Current level 
exceeds 3A, the overload 
indicator turns red and CH3 
operation mode switches from 
Constant Voltage to Constant 
Current. 
 
 
Note: “overload” in this case does not mean an 
abnormal operation.