SAFETY INSTRUCTIONS
5
SAFETY INSTRUCTIONS
This chapter contains important safety instructions
that you must follow when operating the GPD-
3303 series and when keeping it in storage. Read
the following before any operation to insure your
safety and to keep the best condition for the GPD-
3303 series.
Safety Symbols
These safety symbols may appear in this manual or on the GPD-3303
series.
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the GPD-3303 series or
to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
GPD-3303 Series User Manual
6
Safety Guidelines
General
Guidelines
CAUTION
Do not place any heavy object on the GPD-3303
series.
Avoid severe impacts or rough handling that
leads to damaging the GPD-3303 series.
Do not discharge static electricity to the GPD-
3303 series.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at circuits directly
connected to Mains (see note below).
Do not disassemble the GPD-3303 series unless
you are qualified as service personnel.
(Measurement categories) EN 61010-1:2001 specifies the
measurement categories and their requirements as follows. The
GPD-3303 series falls under category I.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
Measurement category I is for measurements performed on
circuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage: 100V/120V/220V/230V
±10%, 50/60Hz
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid
electrical shock.
Fuse
WARNING
Fuse type: 100V/120V: T6.3A/250V,
220V/230V: T3.15A/250V
Make sure the correct type of fuse is installed
before power up.