OPERATION 
29 
 
 
 Voltage level 
Double the reading on the CH1 
Voltage meter. In the above 
case, the actual output is 20.0 x 2 
= 40.0V. 
 Current level 
CH1 meter reading shows the 
output Current. In the above 
case, 2.000A. (CH2 Current 
control must be in the 
Maximum position=3.0A). 
Tracking series with common terminal 
 
Connection 
 
Output rating 
0~30V/0~3A for CH1 ~ COM 
0~–30V/0~3A for CH2 ~ COM 
    GPD-3303 Series User Manual 
30 
 
1.  Press the SER/INDEP key to 
activate the tracking series 
mode. The key LED turns on.
 
 
2.  Connect the load to the front panel terminals, 
CH1+ & CH2−. Use the CH1 (−) terminal as the 
common line connection. 
 
 
 
Note: this diagram shows non-European terminals. 
 
3.  Press the CH1 key (LED turns 
on) and use the Voltage knob 
to set the master & slave 
output voltage (the same level 
for both channels). 
By default, the Voltage and 
Current knob work in the 
coarse mode. To activate the 
fine mode, press the knob to 
turn the FINE LED on. 
 
(master & slave) 
 
(Fine control) 
 
 
 
•  Coarse: 0.1V or 0.1A @ rotation click 
•  Fine: the smallest digit @ rotation click 
 
4.  Use the Current knob to set 
the master output current. 
 
 
5.  To turn on the output (and 
LED), press the output key.