GPT-9000 Series User Manual 
120 
 
Sets the MANU test number.    
MANU:STEP <NR1> 
MANU:STEP? 
Parameter/ 
Return parameter 
MANU:STEP 100 
Sets the manual test number to 100. 
 
Sets or returns the test name for the selected 
manual test. The test must be in MANU mode 
before this command can be used.  
Note only alphanumeric characters (A-Z, a-z, 0-9) 
and the “_” underscore character can be used to set 
the MANU test name.  
MANU:NAME <string> 
MANU:NAME? 
Parameter/ 
Return parameter 
10 character string. (first character must 
be a letter) 
MANU:NAME test1 
Sets the manual test name to “test1”. 
 
Sets or returns the Ramp Time for the test in 
seconds.  
 
Note: A “TIME ERR” will result  if the Ramp Time 
+ Test Time is ≥ 240 seconds when the HI SET limit 
is over 30mA (ACW function only).