REMOTE CONTROL
129
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value in
M. The test must first be in IR mode before this
command can be used.
Syntax
Query Syntax
MANU:IR:RHISet <NR1>|NULL
MANU:IR:RHISet?
Parameter/
Return parameter
<NR1>
2 ~ 9999
NULL
Sets the HI SET value to high impedance
Example
MANU:IR:RHIS 10.
Sets the IR HI SET resistance to 10 M.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
M. The LO SET value must be less than the HI
SET value. The test must first be in IR mode before
this command can be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR1>
1 ~ 9999
Example
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10M.