EasyManua.ls Logo

GW Instek GPT-9800 - Manu:gb:ref; Manu:gb:zerocheck

GW Instek GPT-9800
156 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
REMOTE CONTROL
133
MANU:GB:REF
Set
Query
Description
Sets or returns the GB reference value in mΩ. The
test must first be in GB mode before this command
can be used.
The GB reference value must be less than the HI
SET value.
Syntax
Query Syntax
MANU:GB:REF <NR2>
MANU:GB:REF?
Parameter/
Return parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:REF 100
Sets the GB reference to 100 m.
MANU:GB:ZEROCHECK
Set
Query
Description
Performs the zero check function. The test must
first be in GB mode and in the Ready Status before
this command can be used.
See page 67 for details on the ZERO function.
Syntax
Query Syntax
MANU:GB:ZEROCHECK {ON|OFF}
MANU:GB:ZEROCHECK?
Parameter/
Return parameter
<ON>
Zero function is active.
<OFF>
Zero function is not active.
Example
MANU:GB:ZEROCHECK OFF
Activates the ZERO function.

Table of Contents

Related product manuals