GPT-9000 Series User Manual
128
MANU:DCW:REF <NR2>
MANU:DCW:REF?
<NR2>
000.0 ~ 010.9
MANU:DCW:REF 0.01
Sets the DCW reference to 0.01 mA.
MANU:DCW:ARCCurrent
Set
Query
Sets or returns the DCW ARC current value in mA.
ARC must be enabled to set the ARC current. The
test must first be in DCW mode before this
command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
MANU:DCW:ARCCurrent <NR2>
MANU:DCW:ARCCurrent?
<NR2>
000.0 ~ 22.00
MANU:DCW:ARCC 10
Sets the DCW ARC value to 10mA.
MANU:IR:VOLTage
Set
Query
Sets or returns the IR voltage in kV. The test must
first be in IR mode before this command can be
used.
MANU:IR:VOLTage <NR2>
MANU:IR:VOLTage?
<NR2>
0.05 ~ 1 (0.05kV to 1kV: steps of .05)