GPT-9000 Series User Manual
148
MANU Setting Errors
The following error messages or messages may appear on the GPT
screen when configuring the MANU test settings.
For ACW tests. TIME ERR is displayed when HI
SET ≥ 30.00mA~40.00mA and if the RAMP
time and the TEST TIME setting is > 240
seconds.
For DCW tests. OVER 50W is displayed if the
HI SET setting multiplied by the Voltage setting
is greater than 50W.
For ACW, DCW tests. Shown when the current
is set too high.
Voltage is too low. Indicates that the DUT
could be shorted.
For ACW, DCW tests. Voltage is too high.
For GB tests. Voltage is equal to 0. Check to
see that the SENSE H or SOURCE H is not
open.
For IR tests. The voltage is too high or
resistance=0Ω. Check to see whether the DUT
or test lead is shorting.
For GB tests. The resistance is too high or too
low. Check the test lead connections to correct
the error.
For GB tests. Current too low. Indicates that
the SOURCE L or SOURCE H test lead is open
or poorly connected. Test the test lead
connection with the DUT to confirm.
For GB tests. Current is too high.
For GB tests. Resistance = 0. This error
indicates that there is an error with the
measured resistance (0Ω). Perform the zeroing
function again.