PSM-SERIES PROGRAMMABLE POWER SUPPLY 
USER MANUAL 
 
  
 
46
6-4. Adjustment and Calibration   
--Preparation 
1.  30 minutes warm up before calibration. 
2.  Ambient temperature:23±5°C, Humidity: Under RH80%. 
--Output Calibration Steps: 
[Step 1] 
Press [SHIFT][ .] to appear Password input window, input Password (vary 
with different models: PSM-2010= 2010, PSM-3004 = 3004, PSM-6003 = 
6003) by using the number key, press [ENTER] key 
[Step 2]   
When get into Calibrated item selection window, set the cursor to the 
calibrated item by using the knob and press [ENTER]. After the calibrated 
item is selected, proceeding voltage, current and over-voltage calibration 
step by step. 
[Step 3] Voltage Calibration Steps 
Set to [VOLTAGE] by using the knob and press [ENTER] getting into 
voltage calibration procedure. 
PSM-SERIES PROGRAMMABLE POWER SUPPLY 
USER MANUAL 
 
  
 
47
[Step 3.1] 
Firstly, proceed the calibration procedure of LO calibrated point, input 
DMM measurement value, then press [ENTER]. 
 
Note: The DMM selected for the measurement must have the resolution of 
4 digits after the decimal point at least, and take three digits after the 
decimal point to be the effective value and run off the rest. 
[Step 3.2] 
Next, proceed the calibration procedure of MI calibrated point, input 
DMM measurement value, then press [ENTER]. 
Note: The DMM selected for the measurement must have the resolution of 
4 digits after the decimal point at least, and take three digits after the 
decimal point to be the effective value and run off the rest. 
[Step 3.3] 
Finally, proceed the calibration procedure of HI calibrated point, input 
DMM measurement value, then press [ENTER]. 
 
Note: The DMM selected for the measurement must have the resolution of 
4 digits after the decimal point at least, and take three digits after the 
decimal point to be the effective value and run off the rest. 
Now, switch to SAVE by using the knob and press [ENTER] to leave the 
mode with storing.    Also can press [SHIFT][ .] to leave the mode without 
storing. 
【Step 4】