13
DEUTSCHENGLISH FRANÇAIS ITALIANO ESPAÑOL NEDERLANDS
PORTUGUÊS
SVENSKA
DEUTSCHENGLISHFRANÇAISITALIANOESPAÑOLNEDERLANDS
PORTUGUÊS
SVENSKA
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland ‑ HS‑Doc. No. 1500.7221019.04030 / 2024 – 04
6.6 Image review mode
• By pressing the menu key (75) you can switch from capture to review mode and
vice-versa.
• In the review mode all captured images and videos of the session are listed in
chronologic order. With the inner key (74) and outer key (73) of the release mod-
ule RM03 you can scroll through the list. Please note that the images saved on
the connector box CB01 will be deleted after 7 days.
• Deleting images and videos: In the image review mode selected images and
videos can be deleted by pressing the trigger key (76).
6.7 Menu
• By pressing the menu key (75) for two seconds you enter the menu. In the menu
you can initiate a white balancing, enter the settings and find the 'About' dia-
logue by selecting the keys (73), (74) and (76).
• Under 'Settings' you'll find 'Audio', 'Capture Mode', 'Aperture Mode', 'Focus
Mode' and 'End Examination Mode'.
◦ 'Audio' – Determines if an acoustic confirmation of the key buttons of the
RM03 is provided.
▪ 'Audio ON' – Acoustic confirmation is activated.
▪ 'Audio OFF' – Acoustic confirmation is deactivated.
◦ 'Capture Mode' – Defines how the images are selected during the capturing
process.
▪ 'Best Image' – Selects the best image of the last 30 frames for you. This
increases the overall success rate of good pictures, however, in certain
situations it may theoretically happen that a picture is selected that you
would have discarded.
▪ 'Newest Image' – Selects the most recent image at the exact moment
the release key (76) is pressed.
◦ 'Aperture Mode' – Determines which mode is set as default at the start of the
examination.
▪ 'Aperture Auto' – Sets auto aperture control as default at the start of the
examination.
▪ 'Aperture Manual' – Sets manual aperture control as default at the start
of the examination.
◦ 'Focus Mode' – Hides the user interface in the image capturing screen.
▪ 'Focus Mode Direct Capture' – Hides the user interface by maintaining
the regular workflow.
▪ 'Focus Mode Freeze Capture' – Hides the user interface and displays a
freeze preview of the captured image (video recording is disabled when
'Focus Mode Freeze Capture' is selected).
▪ 'Focus Mode OFF' – Enables the user interface in the image capturing
screen.
◦ 'End Examination Mode' – Offers the possibility of summarizing recordings
to examinations, initiated by switching the beam splitter switch and confirm-
ing the end of the current examination.
▪ 'End Examination Mode ON' – Activates the mode. When enabled, an
'End of Examination' prompt appears when the beam splitter switch is
operated. When confirmed, the system will store all images and videos
taken during the examination in a corresponding folder on the connector
box CB01. Simultaneously it will start a new examination session. (The
'End of Examination?' prompt can also be initiated by pressing the inner
key (74) and outer key (73) simultaneously when you are in the preview
mode).
▪ 'End Examination Mode OFF' – Deactivates the mode.