9 Maintenance
WARNING!
Do not modify this device without authorization of the manufacturer.
Installation and repairs may only be performed by trained special-
ists.
Contact your Haag-Streit representative for installation, repairs and
modification work on the system. The contact details are available at
www.haag-streit.com.
Only original Haag-Streit spare parts may be used.
9.1 Device inspection
To check the correct function, proceed as follows:
Insert the test rod into the radial movement bearing, whilst at the same time
aligning the surface to the microscope at a right angle.
Set the slit length to 8 or 14 mm.
Set the illumination intensity to 50%.
Set the magnification in the microscope to max.
Set the eyepieces in such a way that the test rod is in sharp focus. In doing so,
turn the ocular from the (+) to the (-) side.
Switch on the camera.
For all magnifications, the structure of the test rod in the ocular and in the cam-
era image must be shown in sharp focus.
Turn on the illumination of the 0.2 mm diaphragm.
Turn on the crosshairs in EyeSuite; the cross must be inside the illuminated dot.
9.2 Servicing
To guarantee a long service life, the device must be cleaned weekly as described
and protected with the dust cover when not in use. We recommend having the
device inspected once a year by an authorized service technician.
9.3 Cleaning and disinfection
The Haag-Streit slit lamps and their accessories can, if required, be carefully wiped
down with ready-for-use disposable 70% ethanol disinfectant wipes. Surface-friendly
disinfectants (containing aldehyde or aldehyde-free) are also permitted, such as
Kohrsolin FF.
WARNING!
The preparation instructions provided do not apply to tonometer
measuring prisms.
Tonometer measuring prisms must be prepared in accordance with
a different manual
Too strong or aggressive disinfectants or cleaning liquids e.g. hydro-
gen peroxide will damage the finish and coating of the device.
Do not use sprays
Observe the manufacturer's safety instructions
Do not use any cloths that drip
Wring out any soaked cloths before use when necessary
Ensure that no liquid penetrates the device
Comply with the stipulated exposure time
Clean optical surfaces after disinfection with a very soft cloth
NOTE!
IP code: IPX0 (device is not protected against liquids)