6.16 Error messages (illumination head)
ERROR 1) 2) 3)
Background illu-
mination
Polarity reversal Slit illumination
Error messages Measures
E1 Incorrect supply polarisation Contact your Haag-Streit representative. x Red x
E2 Illumination control not recog-
nized
Connect illumination control or replace, if necessary. Red
x Red
E3 Temperature is too high The light sources’ power will be reduced. Normal operation is
ensured once the permissible temperature has been reached.
Red, flashing x Red, flashing
E4 No communication between
power supply and illumination
Contact your Haag-Streit representative. Red, flashing 2 × x Red, flashing 2 ×
E6 General error Send PS-LED to the appropriate service branch. Red, flashing 4 × x Red, flashing 4 ×
6.17 Error messages release module RM03
ERROR Error messages Measures Operational control LED (67)
E10 General error Send device to the appropriate service branch. Red
E12 Communication error Contact your Haag-Streit representative. Red, flashing 2 ×
E14 Hardware version error Contact your Haag-Streit representative. Red, flashing 4 ×