4. When the upgrade is complete, the display shows TRANSFER COMPLETE. Remove the SD
card.
5. Restart the instrument for the upgrade to take effect.
Calibration
N O T I C E
The manufacturer recommends calibration of the analyzer after 1 day of operation to allow all system components
to stabilize.
The automatic calibration uses the installed known standards to calibrate the analyzer. Calibration
can be done manually or scheduled to occur automatically. Use the calibration menu to view
calibration data, start or cancel the auto calibration, override the auto cal settings, calibrate manually
or to reset to the default calibration.
1. Push cal to display the calibration menu.
Option Description
START MANUAL CAL Refer to Start a calibration manually on page 20.
SET AUTO CALIBRATION Refer to Schedule automatic calibrations on page 20.
CALIBRATION DATA Displays data from the last calibration and the due date and time of the next
scheduled calibration.
OVERRIDE CALIBRATION Enter new slope and/or zero value (reagent blank). When OVERRIDE
CALIBRATION is selected, the automatic calibration is disabled for the selected
option.
OUTPUT CALIBRATION Select a 4-20 mA output and enter the output values to be sent.
RESET DEFAULT CAL Reset calibration data to default values and disables automatic calibration. On
completion, perform a new calibration.
Schedule automatic calibrations
1. Push cal, then select SET AUTO CALIBRATION.
2. Select SLOPE CAL or ZERO CAL.
3. Select ACTIVATE AUTO CAL>YES.
4. Select STD SOLUTION and enter the standard value in ppb(not applicable for ZERO CAL).
5. Select a schedule option for calibration.
Option Description
TIME BASE Sets the interval between calibrations. Options: DAYS or HOURS.
WEEK DAY Selects the day or days of the week for calibration when TIME BASE is set to DAY.
TIME Sets the time of day for calibration when the TIME BASE is set to DAY.
SET INTERVAL Sets the interval between automatic calibrations in hours when the TIME BASE is set to
HOURS .
Start a calibration manually
1. Push cal, then select START MANUAL CAL.
2. The type of calibration is displayed. Push enter to start the calibration.
20
English

Other manuals for Hach 5500sc SiO2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 5500sc SiO2 and is the answer not in the manual?

Hach 5500sc SiO2 Specifications

General IconGeneral
Power Requirements100 - 240 VAC, 50/60 Hz
Resolution0.001 mg/L
Light SourceLED
DetectorPhotodiode
Sample Flow Rate200 to 500 mL/min
Operating Temperature5 to 45 °C
Sample Temperature5 to 45 °C
Outputs4-20 mA
EnclosureNEMA 4X/IP66
MethodHeteropoly blue method