12
Installation
3.6 Cell Compartments, Cell Adapters, Light Shield, and Protective Cover
3.6.1 Cell Compartments and Adapters
The DR 2800 has two cell compartments (Figure 3). Cell
Compartment #2 uses adapters (Figure 4) to accommodate
different cell types. Only one cell type at a time can be used for a
measurement.
Cell Compartment #1
• 13-mm TNTplus™ or 16-mm TNT round cells
Note: Cell Compartment #1 contains a barcode reader for
TNTplus™ vials.
Cell Compartment #2
• 1-inch square or 50-mm rectangular cells (can be inserted
directly into the cell compartment without using an adapter).
• Adapter A: 10-mm square cells
• Adapter B: Pour-Thru and multi-path cells
Note: Pour-Thru cell must be used with Adapter B, not Adapter C.
• Adapter C: One-inch round cells and AccuVac
®
Ampuls
Note: One-inch round cells and AccuVac Ampuls must be used with
Adapter C, not Adapter B.
Figure 3 Cell Compartments
1 Cell Compartment #1 2 Cell Compartment #2