EasyManuals Logo

Hach LDO101 User Manual

Hach LDO101
18 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
Figure 3 Probe exploded view
1 Shroud 7 Probe lens
2 Locking ridges (8x) 8 Cap seal
3 Locking rib 9 Temperature sensor
4 Locking ring 10 Locking groove
5 LDO sensor cap 11 Locking ribs (4x)
6 O-ring
7.2 Replace the sensor cap and iButton
Replace the LDO sensor cap after 365 days, or more frequently if the cap becomes damaged or
dirty. The LDO sensor cap and iButton operate together and must be replaced at the same time.
Keep the shroud on the probe during use to prevent damage to the LDO sensor cap. Refer to the
instructions that are supplied with the LDO sensor cap replacement kit.
7.3 Remove or install the shroud
Remove the shroud during calibration and maintenance. Refer to Figure 4. Keep the shroud installed
on the probes during sample measurements to prevent damage to the LDO sensor cap. Refer to
Figure 5.
English
9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach LDO101 and is the answer not in the manual?

Hach LDO101 Specifications

General IconGeneral
BrandHach
ModelLDO101
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals