Use the filter scan probe
N O T I C E
The filter scan probe function applies to 1 CFM and 50 LPM units only.
1. On the Counter Navigation screen, push SAMPLE.
2. In the Test screen, push the FILTER icon.
3. To start the test, push START FILTER PROBE TEST.
4. To end the test, push STOP FILTER PROBE TEST.
5. Push PRINTER to generate a brief report of the last completed test.
Manage sample batch identification
Batch IDs are used to label different test runs. Labels can be text or numbers. The Batch ID is
shown:
• On the main sample screen in the settings
• In the System Messages of the Historical screen
• In printed reports, FTP exports and all USB exports
Enter or change a Batch ID
To enter or change a Batch ID:
1. On the Counter Navigation screen, push SAMPLE.
2. Select the Settings tab.
3. Select BATCH ID.
4. Enter the Batch ID as a text, numerical or alphanumerical string. The Batch ID can use up to
29 characters.
5. Push ENTER.
Disable a Batch ID
Disable the Batch ID to stop it from showing in data reports.
1. On the Counter Navigation screen, push SAMPLE.
2. Select the Settings tab.
3. Select BATCH ID.
4. Delete the Batch ID text.
5. Push ENTER. The default Batch ID value will show, but will not be reported.
Set or clear the Batch ID
At the start of a sample test push BATCH ID. The user can select:
• Continue—Keep the Batch ID the same.
• Clear ID—No Batch ID is recorded with the sample cycle.
• Edit ID—Change the Batch ID for the current sample cycle.
The Batch ID can also be cleared when the sample cycle is complete.
How to use the Test and Report Wizard
About standard sampling protocols
Use the Test and Report Wizard for groups, areas and locations that require sample
measurements and reports based on specific standards and regulatory guidelines. The 3400-series