Table 2 Beverage (continued)
Application Sensor Membrane Cartridge Protection Cap
Wort with ORBISPHERE
3650 Portable and special flow
cell 32007W.xxx
GA2400-S00 29552A-A 33051-S0
ORBISPHERE 6110 Total
package analyzer (TPA)
GA2400-S00T 2956A-A 33051-ST
De-aerated water GA2400-S00 2956A-A or 2952A-A 33051-S0
33051-SG (if exposed to
CIP)
Table 3 Pure water applications (Power - Electronics)
Application Sensor Membrane Cartridge Protection Cap
On-line dissolved oxygen traces in pure water GA2400-S00 2956A-A 33051-S0
Dissolved oxygen traces with 3655 Portable GA2400-S00 2956A-A 33051-S0
Table 4 ATEX
Sensor Membrane Cartridge Protection Cap
GA2800-SVS 2956A-A or 29552A-A 33051-SG
GA2800--SKS 2956A-A or 29552A-A 33051-SG
GA2800--HVS 2956A-A or 29552A-A 33051-H0
GA2800--HKS 2956A-A or 29552A-A 33051-H0
2.3 Sensor membrane specifications
2.3.1 Oxygen sensors (Table 1)
Table 5 Membrane specifications - Oxygen sensors (1)
Specification 2956A-A 2958A-A 29552A-A 2952A-A
Recommended applications Corrosion control,
De-aerated water
Beverage,
Lab. applications
In line wort,
Air/O
2
injection,
Sewage treatment
Corrosion control,
In line beverage,
De-aerated water
Material PFA Tefzel
®
2
PTFE Tefzel
®
Thickness [μm] 25 12.5 50 25
Calibration gas Air Air Air Air / Pure O
2
Dissolved measurement range 0 ppb to 20 ppm 0 ppb to 40 ppm 0 ppb to 80 ppm 0 ppb to 80 ppm
Gaseous measurement range 0 Pa to 50 kPa 0 Pa to 100 kPa 0 Pa to 200 kPa 0 Pa to 200 kPa
Accuracy The greater of
±1% of reading
or ± 0.1 ppb
3
,
or ± 1 ppb
4
,
or ± 0.25 Pa
The greater of
±1% of reading
or ± 1 ppb,
or ± 2 Pa
The greater of
±1% of reading
or ± 2 ppb,
or ± 5 Pa
The greater of
±1% of reading
or ± 2 ppb,
or ± 5 Pa
2
Tefzel is a trademark of DuPont Corporation.
3
Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments
4
Accuracy is ± 1 ppb for 366x and 3650 instruments
4 English