EasyManua.ls Logo

Hach POCKET COLORIMETER II - Exiting the Calibration Routine; Deleting Calibration Points

Hach POCKET COLORIMETER II
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2—24
User-Entered Calibration, continued
Note: When a factory calibration curve has been edited, the “calibration adjust” icon will
appear in the display.
Exiting the Calibration Routine
Exit the calibration routine by pressing the MENU key to return to measurement
mode. The instrument uses the last completed user-entered calibration or the
factory calibration if no user-entered calibration has been completed.
Deleting Calibration Points
1. Select the range containing user-entered calibration points. See Switching
Ranges on page 2—4.
2. Press and hold the
MENU key until “USER”, then “CAL” appears. Press
READ/ENTER.
Note: Calibration points can also be deleted in Edit mode.
3. Press ZERO/SCROLL to select the point to delete (e.g., S0 or S1 or S2). Press
READ/ENTER.
4. The left digit will flash. Press
ZERO/SCROLL until “dEL” appears. (“dEL” will
appear after the numeral 9.)

Table of Contents

Other manuals for Hach POCKET COLORIMETER II

Related product manuals