2—22
User-Entered Calibration, continued
10. When all the calibration data has been entered, press MENU twice to return to
the measurement mode.
Editing a User-entered or Factory Calibration Curve
1. Press the MENU key and hold it down until the display shows “USER”, followed
by “CAL”. Press
ZERO/SCROLL until EDIT appears.
2. Press the
READ/ENTER key to enter Edit mode. In factory-calibrated meters,
“S0” will appear in the display.
Note: When editing a factory-calibrated meter or range, RES (resolution) cannot be changed.
Note: When RES or S0 appears in the display, press
ZERO/SCROLL
to quickly scroll to the
data to be edited.
3. In uncalibrated meters or in Abs range, “RES” will appear. Press ZERO/SCROLL
to review the current resolution. Press
ZERO/SCROLL again to accept the
displayed resolution. To change the resolution (decimal placement), press
READ/ENTER. Press ZERO/SCROLL to select the new resolution, then press
READ/ENTER to accept. “S0” will appear on the display.
4. Press
READ/ENTER. The current concentration value for S0 will appear on the
display.