1—27
Bromine, Total, HR (0.2 to 10.0 mg/L Br
2
)
Method 8016
For water, wastewater, and seawater
DPD Method* (PermaChem
®
Powder Pillows)
Measuring Hints
If the bromine concentration is typically less than 4 mg/L, use the low range
procedure.
Analyze samples immediately. Do not use plastic containers to collect samples.
Note:
The Pocket Colorimeter II is designed to measure solutions contained in sample cells.
DO NOT
dip the meter in the sample or pour the sample directly into the cell holder.
* Adapted from Standard Methods for Examination of Water and Wastewater.