EasyManua.ls Logo

Hach POCKET COLORIMETER II - Page 16

Hach POCKET COLORIMETER II
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1—16
Bromine, Total, LR, continued
Bromine, Total, Low Range (using Powder Pillows)
1.Fill a 10-mL sample cell
sample (the blank). Cap.
Note:
Samples must be
analyzed immediately and
cannot be preserved for later
analysis.
2.Press the POWER key to
turn the meter on.
The arrow should indicate
the low range channel (LR).
Note:
See page 2—4 for
information on selecting the
correct range channel.
3.Remove the instrument
cap. Place the blank in the
cell holder. Cover the blank
with the instrument cap.
Note:
Wipe excess liquid and
fingerprints from the sample
cells.

Table of Contents

Other manuals for Hach POCKET COLORIMETER II

Related product manuals