Chlorine Dioxide, continued
4.Press ZERO/SCROLL.
The display will show
“- - - -” then “0.00”.
Remove the blank from the
cell holder.
5.Collect 40 mL of sample
in a 50- or 100-mL beaker.
Add 16 drops of Glycine
Reagent to the 40 mL of
sample. Swirl gently to mix.
Continue with step 6
immediately.
6.Fill a DPD Free Chlorine
Reagent AccuVac Ampul
with the pretreated sample in
the beaker.
Note: Keep the tip immersed
until the ampule fills
completely.