EasyManuals Logo

Hach POCKET COLORIMETER II User Manual

Hach POCKET COLORIMETER II
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
2—14
Standard Calibration Adjust, continued
7. Press ZERO/SCROLL to access the Edit function, then press READ/ENTER to
begin editing. The digit to be edited will flash. Use the
ZERO/SCROLL key to
change the entry, then press
READ/ENTER to accept and advance to the
next digit.
When the last digit is entered, press
READ/ENTER and the meter will adjust the
curve to the value entered. The meter will return to measurement mode and the
Calibration Adjusted icon will appear in the display window.
To turn off Standard Calibration Adjust (SCA):
1. Press
MENU.
2. Press
ZERO/SCROLL until “SCA” appears in the display.
3. Press
READ/ENTER, then press ZERO/SCROLL until “Off” appears in the display.
4. Press
READ/ENTER to turn off SCA.
Note: Perform another standard calibration adjust to turn SCA on again.
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration. See Retrieving the
Factory Calibration on page 2—25.

Table of Contents

Other manuals for Hach POCKET COLORIMETER II

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach POCKET COLORIMETER II and is the answer not in the manual?

Hach POCKET COLORIMETER II Specifications

General IconGeneral
Measurement MethodColorimetric
DisplayLCD
Operating Temperature0 to 50 °C
Storage Temperature-40 to 60 °C
Waterproof RatingIP67
Weight0.23 kg
Light SourceLED
RangeVaries by parameter
WavelengthVaries by parameter
Battery LifeApproximately 2000 tests

Related product manuals