Table of Contents
1 Specifications on page 3
2 General information on page 4
3 Installation on page 7
4 Startup on page 16
5 User navigation on page 16
6 Operation on page 16
7 Maintenance on page 20
8 Troubleshooting on page 23
9 Consumables and replacement parts on page 27
Section 1 Specifications
Specifications are subject to change without notice.
Refer to the SC4500 Controller documentation for the SC4500 Controller specifications.
Specification Details
Dimensions Sensor with membrane—108 mm x ∅ 30 mm (4.25 x ∅ 1.18 inches)
Flow chamber—68 x ∅ 50 mm (2.68 x ∅ 1.97 inches)
Weight Sensor with membrane—0.066 kg (0.15 lbs)
Flow chamber—0.74 kg (1.63 lbs)
Material Body and membrane holder: Noryl; body with optional immersion: stainless steel
316L; electrodes: gold cathode and silver anode; membrane: PFA; flow
chamber: stainless steel 316L
Pollution degree 2
Overvoltage category II
Protection class I, connected to protective earth
Environmental conditions Indoor and outdoor use
Altitude 2000 m (6562 ft) maximum
Operating temperature -20 to 60 °C (-4 to 140 °F)
Humidity 0—95% relative humidity, non-condensing
Sample Flow rate: 66 to 166 mL/min (4 to 10 L/h)
Temperature: -20 to 60 °C (-4 to 120 °F)
Pressure: atmospheric pressure
Calibration Zero—Electrically or with oxygen-free water
Slope—In air or against a laboratory measurement
Measurement range 0 to 2000 ppb (0 to 2 ppm); 0 to 45 °C (32 to 113 °F)
Sensitivity < 0.5 ppb
Repeatability ± 0.5 ppb or ± 5 % of measurement (the larger value)
Reproducibility ± 0.5 ppb or ± 2 % of measurement (the larger value)
Detection limit ≤ 1 ppb
Response time Change of 1 to 40 ppb: < 30 seconds
Temperature compensation Automatic in the range 0 to 45 °C (32 to 113 °F)
Analog outputs (optional)
1
Five 0–20 mA (or 4-20 mA) analog outputs on each analog output module
2
1
Dependent on controller configuration.
2
Refer to the module documentation for additional information.
Note: Install only one module in one of the available slots.
English 3