Transmitter 9135 - pH/redox measurement
Page 40
S/MEASURE Menu
ELECTRODE menu allow to set up the electrode
model.
TEMP. COMP. menu allow to set up the
temperature compensation.
ELECTRODE
Choice of the electrode.
TYPE : • Glass
• Antimony
• Redox
• Other
ISO Pt : XX,XX pH
The isothermal point corresponds to
the pH value which does not vary
according to the temperature.
SLOPE : XX %
Indication of the electrode sensitivity
in % of the theoric value (59,16
mV/pH at 25°C).
Uiso : XXXX mV
Uiso is the potential of the isothermal
point.
TEMP. COMP. :
MEASURE
Select
ELECTRODE
TYPE :
ISO Pt : 1,5 pH
SLOPE : 95 %
Uiso : 200 mV
ELECTRODE
Redox