Transmitter 9135 - pH/redox measurement
Page 42
S/mA OUTPUTS Menu
This menu allow to adjust analogue outputs.
OUTPUT 1/2
− AFFECT : choice if the analogue output
allocated to measure or
temperature.
• pH/mV
• °C/°F
− TYPE : choice of the analogue output type.
• 0/20 mA
• 4/20 mA
− LOWER : lower limit value programming.
− UPPER : upper limit value programming.
OUTPUT 1
OUTPUT 2
SPECIAL PROG.
TEST
mA OUTPUTS
Select
AFFECT. :
TYPE : 0/20 mA
LOWER : 10.0 mS
UPPER : 20.0 mS
OUTPUT 1/2
Select
pH