f. Select Controller > Connectivity > HCNB > HCNB. Select On. Push OK to restart the
controller.
Note: Wait until the controller restarts. Do not touch the display or go to other menus or an error message
will show.
2. Set the Controller 2 configuration the same as Controller 1.
3. Set The Controller 3 configuration as follows:
Note: Make sure to connect the Ethernet LAN cable to LAN port 1 and industrial Ethernet cable to IEP port 1.
a. Push the main menu icon, then select Controller > Connectivity > LAN > Ethernet ports.
Select Split or Mix IEP.
b. Go to Controller > Connectivity > HCNB > HCNB. Make sure that HCNB is set to Off.
c. Select Controller > Connectivity > HCNB > Interface. Select ALL.
d. Select Controller > Connectivity > HCNB > Auto detection. Select On.
e. Select Controller > Connectivity > HCNB > HCNB. Select On. Push OK to restart the
controller.
Note: Wait until the controller restarts. Do not touch the display or go to other menus or an error message
will show.
English 11

Other manuals for Hach SC4500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach SC4500 and is the answer not in the manual?

Hach SC4500 Specifications

General IconGeneral
Power Supply100-240 VAC, 50/60 Hz
Operating Temperature-20 to 60 °C (-4 to 140 °F)
EnclosurePolycarbonate
ConnectivityEthernet, USB
Enclosure RatingIP66, NEMA 4X
CommunicationHART, Profibus
Measurement ParameterspH, ORP, conductivity, dissolved oxygen, turbidity