EasyManua.ls Logo

Hach Sigma 1600 - Section 1 Overview; 1.1 Measurement Capabilities

Hach Sigma 1600
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 9
6507Overview.fm Overview
Section 1 Overview
The Sigma 1600 Automatic Liquid Sampler automatically collects wastewater
samples and preserves the samples using refrigeration. The sampler
operates on 120 V ac, 60 Hz. The sampler cabinet is constructed entirely of
non-corrosive fiberglass reinforced plastic. The refrigerator is equipped with a
1
/6 horsepower compressor with a fan-cooled condenser, front ventilation,
silver brazed connections, and adjustable thermostat. Refer to Figure 1 .
Figure 1 Sigma 1600 Automatic Liquid Sampler
1.1 Measurement Capabilities
The sampler collects samples on a timed-cycle basis or in proportion to flow
rate when used in conjunction with a flow meter. The flow through chamber
(Figure 2) receives a continuous flow of wastewater supplied by a pump or
gravity flow. Gravity flows the wastewater out of the sampler chamber through
a 3 in. diameter outlet. To prevent suspended solids from settling, the flow rate
through the sampler chamber can be 5–30 gallons per minute. The rotating
cam operated dipper arm cycle takes a repetitive 25 mL sample from the
chamber’s wastewater flow.