7
Specifications
Figure 2 Maximum ambient temperature vs. probe load
Table 2 sc1000 controller capacity
Controller operating temperature (°C) Power available (watts) Controller capacity
40 55 2 SS7 sc plus 15 watts for other devices
50 35 2 SS7 sc plus 11 watts for other devices
55 25 1 SS7 sc plus 13 watts for other devices
38
40
44
48
52
56
42
46
50
54
58
Probe Load (watts)
sc 1000 Ambient T (°C)
Temp
20
22 24
26 28
30 32 34 36 38
40
42
44
46 48 50 52
54
56
58
Table 3 sc1000 component power consumption
Component Power consumption (watts)
Display module 10
Current output card 2.5 maximum load
Current input card 1.5
Relais card 1
Fieldbus module (Profibus) 2.5
Fieldbus module (Modbus) 0.5