17
Installation
turbidimeter sensor must be mounted within six feet of the
controller unless an extension cable is used. Maximum cable length
is 9.6 m (31.5 ft).
3.3.3 Mounting the SS7 sc or SS7 sc-HST
1. To ensure proper performance, the sample unit must be level
(Figure 9 on page 19). Use a small level across the top opening
of the turbidimeter body to verify that the instrument does not
slope left-to-right or front-to-back.
2. Use one rubber washer (supplied) at each wall mounting block
between the block and the wall. Mounting blocks are secured to
the four corners of the sample unit to facilitate wall mounting
without affecting the integrity of the enclosure protection.
3. Attach the sample unit to the wall with four customer-supplied
mounting bolts.