19
Installation
3.3.4 Installing the optional heat exchanger
An optional heat exchanger (Cat. No. 48551-00) is available for the
SS7 sc-HST (Figure 10 on page 20). The heat exchanger reduces
sample temperatures that exceed the temperature requirements of
the instrument. It can reduce sample temperatures of up to 100 °C
but is not suitable for steam or super-heated water. A source of
cooling water is required. The heat exchanger is made of 316
stainless steel and has ¾” MNPT pipe connections. The large
plumbing connections help eliminate clogging.
Allow adequate space below and to the right (latch) side of the
sample unit to make hydraulic connections.
See Figure 10 on page 20 for installation dimensions.
See Figure 12 on page 22 for heat exchanger connections.
Figure 9 Instrument leveling
1 Level