Q: What is the product warranty?
A: The warranty is 2 years for both meter and probe.
Q: What is the meter’s water resistance rating?
A: The meter is rated IP55 and cannot be immersed.
Q: Can the meter be connected to a computer for downloading data?
A: No, the meter does not have a way to connect to a storage device or a computer.
Q: What do I do if Hach changes the software version of the meter?
A: The meter will have to be sent to the appropriate service center for the software update.
Q: Is there any precaution to switching curves while still taking a continuous measurement?
A: When switching curves during a measurement, the next measurement is done with the new
calibration curve. The respective data set is stored respectively. Data sets consist of Calibration
curve name/Meas. Value/date and time. If an individual measurement site at the plant is
assigned to one calibration curve, data may easily be identified. If you have two measurement
points using the same calibration curve, you may distinguish the measurement sites only by the
time the measurement was taken. In the field this should work fine since measurements are
done in a predefined order.
Q: How does the DELETE function work under MEMORY?
A: This function deletes memorized measurement data that were performed within the course
of a calibration. In case you have not assigned concentrations or turbidity values (points marked
with *), the calibration measurement can be deleted and retaken.
Q: Is setting a calibration point to zero, the only way to remove the point from a calibration curve?